Error Messages - Microlife CARDIO+ Instruction Manual

For accurate blood pressure measurement
Table of Contents

Advertisement

Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g., «ERR 3», is displayed.
Error
Description
«ERR 1»
Signal too weak
«ERR 2»
Arm movement
«ERR 3»
No pressure in
the cuff
Potential cause and remedy
The pulse signals on the cuff are
too weak. Re-position the cuff and
repeat the measurement.*
During the measurement, error
signals were detected by the cuff,
caused, for instance, by movement
or muscle tension. Repeat the
measurement, keeping your arm still.
Adequate pressure cannot be
generated in the cuff. A leak may
have occurred. Check that the
cuff is correctly connected and
not too loose. Recharge the
batteries if necessary. Repeat the
measurement.
«ERR 5»
Abnormal result
«ERR 6»
MAM mode error
Low battery
* Consult your doctor, if this or any other problem occurs repeatedly.
The measuring signals are
inaccurate and no result, therefore,
can be displayed. Read through
the checklist for performing reliable
measurements and then repeat the
measurement.*
There were errors during the
measurement in MAM mode,
making it impossible to obtain
a final result. Read through the
checklist for performing reliable
measurements and then repeat the
measurement.*
The battery power in the Microlife
Cardio+ Blood Pressure Monitor
Dock is too low. Charge the dock
before taking any measurements.
12
EN