4.4.2.1

Output signal hardware modification 4 to 20 and 0 to 20 mA

To modify output signal from 4 to 20 and 0 to 20 mA make following hardware modification on Analog
In put/Output board:
• Identify Jumper IP1 and JP2 on board (see figure 11);
• Remove JP1 (cut copper, this jumper is factory-made by default);
• Install Jumper in JP2 position.
JP1 copper to be cut
JP1
JP2
Fig. 11 - Jumper position for 4–20 to 0–20 mA output signal modification