Alarms Page - ABB AW400 Series Instruction Manual

Chlorine transmitter
Table of Contents

Advertisement

9.1.3

Alarms page

It is accessed by pressing the Enter key in display mode.
If no alarm or no abnormal condition is present the display will show nothing.
The following messages are displayed with the indicated meaning:
A1L = low alarm on channel 1
A2L = low alarm on channel 2
A3L = low alarm on channel 3
A1H = high alarm on channel 1
A2H = high alarm on channel 2
A3H = high alarm on channel 3
=∂= = the value set for delta has been overcome
AVE = the instrument is computing the average of two measures (same parameter) and the averaged value is
indicated on the first line of the display instead of the measure of the first channel
Stop = When CCI in OR parameter is set to YES it means that output(s) is(are) frozen because either CCI1 or
CCI2 or both is(are) closed.
Run = When CCI in OR parameter is set to YES it means that output(s) is(are) working (not frozen) because
CCI1 and CCI2 are both open.