Output Setting - ABB AW400 Series Instruction Manual

Chlorine transmitter
Table of Contents

Advertisement

4.4.2

Output setting

The Output Settings menu allows to set current output (0–20 or 4–20 mA), zero (Out Zero) and full scale (Out
Max) values, in engineering units. Out Zero value corresponds to 0 mA or 4 mA (according to the output
chosen) and the Out Full Scale value to 20 mA.
WARNING!
(See Section 4.4 to get here)
Menu
Output setting
Output Settings
Channel 1
Channel 1
0-20 / 4-20 mA
Cannel 1
Out Zero
Channel 1
Out Span
In the following table the default values of Out Zero and Out Max are presented. Minimum span and
maximum ranges are also presented.
Parameter
Unit
pH
pH
mV
mV
O3
ppm
Cl
ppm
CD
ppm
T
°C
mA
mA
Set 0–20 mA or 4–20 mA accordingly. To modify output signal change Jumper JP1-
JP2 as shown in Par. 4.4.2.1 – Fig. 11
Output Settings
Channel 2
Channel 2
0-20 / 4-20 mA
Channel 2
Out Zero
Channel 2
Out Span
Only when Channel 2
pc board is installed
Out Zero
Out Max
2.00
12.00
-500
+500
0.00
1.00
0.00
1.00
0.00
1.00
-10.0
100.0
4.00
20.0
Output Settings
Channel 3
Channel 3
0-20 / 4-20 mA
Channel 3
Out Zero
Channel 3
Out Span
Only when Channel 3
pc board is installed
Minimum Span Maximum
1.00
100
0. 25
0.25
0.25
5
2
range
0.00–14.00
-1500–+1500
0.00–10.00
0.00–10.00
0.00–10.00
-20–+100
(0) 4–20.0
31