3.2 Calibration Overview

The LDT-5525 Temperature Controller should be calibrated every 12 months or whenever performance verification
indicates that calibration is necessary.
All calibrations can be done with the case closed. The instrument is calibrated by changing the internally stored
digital calibration constants.

3.2.1 Recommended Equipment

Recommended test equipment for calibrating the LDT-5525 Temperature Controller is listed in Table 6.1. Equipment
other than that shown in the table may be used if the specifications meet or exceed those listed.
Description
Mfg./Model
DMM
HP 3457A
Resistors
Metal Film
High-Power
RECOMMENDED TEST EQUIPMENT
Specification
DC Amps (@ 1.0 A): ±0.02 %
Resistance (@ 10 Ω): 0.02 %
0.1 µA or 0.1 mV resolution
15 KΩ (for ITE calibration)
4 KΩ and 40 KΩ (for
100 µA calibration)
4 KΩ and 400 KΩ
(for 10 µA calibration)
4 KΩ and 10 KΩ (for
LM335 calibration)
10 KΩ and 20 KΩ (for
AD590 calibration)
1 Ω, 20 W, low TCR (for ITE
calibration)
Table 3.1 Recommended Test Equipment
3-2