VI. Temperature Settling Test
1_______
Set the 5525 sensor select switch to "100 uA" (-01-) and set C1 to 0.99, C2 to 2.57.
2_______
Connect the output to the TEC Cal/Test Fixture.
3_______
Set the TEC Cal/Test Fixture to Settling Sensor 1.
4_______
Set the 5525 to T mode, and set the temperature to 2°C.
5_______
Turn the 5525 output on and wait until the output has settled to 2.0°C, plus or minus 0.2°C, usually
about 30 seconds to two minutes.
6_______
Turn the 5525 output off.
7_______
Set the 5525 sensor select switch to "10 uA" (-02-) and set C1 to 0.99, C2 to 2.57.
8_______
Set the TEC Cal/Test Fixture to Settling Sensor 2.
9_______
Set the 5525 to T mode, and set the temperature to 10°C.
10______
Turn the 5525 output on and wait until the output has settled to 10.0°C, plus or minus 0.2°C, usually
about 30 seconds to two minutes.
11______
Turn the 5525 output off.
12______
Set the 5525 sensor select switch to "LM335" (-03-) and set C1 to 0.00, C2 to 1.00.
13______
Set the TEC Cal/Test Fixture to Settling Sensor 3.
14______
Set the 5525 to T mode, and set the temperature to 2°C.
15______
Turn the 5525 output on and wait until the output has settled to 2.0°C, plus or minus 0.2°C, usually
about 30 seconds to two minutes.
16______
Turn the 5525 output off.
17______
Set the 5525 sensor select switch to "AD590" (-04-) and set C1 to 0.00, C2 to 1.00.
18______
Set the TEC Cal/Test Fixture to Settling Sensor 4.
19______
Set the 5525 to T mode, and set the temperature to 2°C.
20______
Turn the 5525 output on and wait until the output has settled to 2.0°C, plus or minus 0.2°C, usually
about 30 seconds to two minutes.
21______
Turn the 5525 output off.