ILX Lightwave LDT-5525 User Manual page 49

Hide thumbs Also See for LDT-5525:
Table of Contents

Advertisement

k Ω
Once the constants A' and B' are determined, the LDT-5525 Temperature Controller is programmed with the following
values of C1, and C2:
C1 = A' * 10
C2 = B' * 10
Computer program
We have included a computer program called STEIN1 that uses a least squares curve fitting routine to determine the
values of C1 and C2. The program is written in IBM's advanced BASICA.
THERMISTOR R/T CURVES
FOR VARYING ROOM TEMP. RESISTANCES
50.00
40.00
30.00
20.00
1 k
2.5 k 5 k
10.00
0.00
-60
-40
0
-20
TEMPERATURE (DEGREES C)
Figure A.1 Thermistor Resistance Versus Temperature
+3
+4
10 k 25 k 50 k
100 k
60
80
20
40
100
A-3
120 140
160

Hide quick links:

Advertisement

Table of Contents
loading