5525 Production Calibration Procedure

LDT-5525 POST BURN-IN CALIBRATION
I. Ite Current Calibration
1_______
Set the sensor select switch to "100 uA" and set C1 to 0.99, C2 to 2.57.
2_______
Connect the output to the TEC Cal/Test Fixture, and set the TEC Cal/Test Fixture for ITE calibration.
3_______
Set the DMM to measure DC volts.
4_______
Hold down both the "MODE" and "DISPLAY" buttons. This will put the 5525 into "ITE CAL" mode.
5_______
Wait until the output has settled to about 3 Amps, usually about 3 seconds.
6_______
Read the voltmeter, and calculate Ite: Ite = Vmeter / Rcal.
7_______
Hold down the "ENBL" button and turn the knob until the display reads the calculated Ite.
8_______
Release the "ENBL" button.
9_______
Allow the Ite current to settle at the new point ( -3 A), again about 3 seconds.
10______
Read the voltmeter, and calculate Ite.
11______
Hold down the "ENBL" button and dial in the new current.
12______
Release the "ENBL" button.
13______
Allow the Ite current to settle at the new point ( +3A).
14______
Calculate the actual Ite current.
15______
Hold down the "ENBL" button and dial in the calculated current.
16______
Release the "ENBL" button.
17______
Allow the Ite current to settle to the new point ( +1 A).
18______
Calculate the actual Ite current.
19______
Hold down "ENBL" and dial in the calculated current.
20______
Release "ENBL".
21______
Allow Ite to settle to the new point (-1 A).
22______
Calculate Ite.
23______
Hold down "ENBL" and dial in the calculated current.
24______
Release "ENBL".
25______
Allow Ite to settle to the new point (-3 A).
26______
Calculate Ite.
27______
Hold in "ENBL" and dial in the calculated current.
28______
Release "ENBL".
29______
Wait for the automatic limit cal procedure to complete - about 10 seconds.
30______
Check the calibration: set the limit to 4A; adjust the Ite control while checking the actual current. Test
Ilim cal: set Ite to 4A, adjust Ilim from 4A down to 0A. Repeat check of Ilim cal for Ite set = -4A.
1
Calibration can be terminated at any time by pressing the "OUTPUT" button. This will stop cal; no constants will
be saved to EEPROM.