LDT-5525 Specifications (continued)
Thermistor range (10 µA):
Thermistor range (100 µA):
Usable Thermistor Range:
Thermistor Resistance
Resolution (10 µA):
Thermistor Resistance
Resolution (100 µA):
Thermistor Resistance
Accuracy:
AD590 Reverse Bias:
LM335 Bias:
User Calibration:
Thermistor:
IC sensor:
Measurement (Display)
Display type:
TE Current Range:
TE Current Resolution:
TE Current Accuracy:
Temperature Range:
Temperature Resolution:
6
Temperature Accuracy
0°C:
20°C:
30°C:
50°C:
____________________________________________________________
6
Accuracy figures quoted are typical for a calibrated 10K thermistor. Accuracy figures are relative to calibration standard and are
dependent on the user-defined configuration of the instrument. Variation from typical value is largely due to uncertainty in
thermistor calibration.
0.0 to 450.0 KΩ
0.00 to 45.00 KΩ
25 to 450,000 Ω, typical
0.1 KΩ
0.01 KΩ
±0.05 % of FS
8 Volts
0.6 mA
Steinhart-Hart equation (2 constants)
2-point
4-digit LED
-4.00 to 4.00 Amps
0.01 Amps
±0.03 Amps
-99.9 to 999.9°C
0.1°C
:
±0.6°C, typical
±0.3°C, typical
±0.4°C, typical
±0.6°C, typical
1-4