ILX Lightwave LDT-5525 User Manual page 42

Hide thumbs Also See for LDT-5525:
Table of Contents

Advertisement

III. 10 uA Sensor Calibration.
1_______
Set C1 to 0.99, C2 to 2.57.
2_______
Set the sensor select switch to "10 uA" (-01-).
3_______
Connect the output of the LDT-5525 to the 5525 Sensor Test Fixture. Set the fixture for SENSOR 2,
5
Load 1
.
4_______
Hold down both the "MODE" and "SENSOR CAL" buttons. This will put the 5525 into "Sensor Cal"
6
mode.
5_______
Allow the sensor measurement circuit to settle, usually about 3 seconds.
6_______
Hold down the "ENBL" button and turn the knob until the display reads the value of the calibrated
resistor, in kΩ (4.0).
7_______
Release the "ENBL" button and wait for the "VIEW SET" light to go out.
8_______
Set the 5525 SENSOR Test Fixture Load switch to position 2
9_______
Allow the sensor measurement circuit to settle.
10______
Hold down the "ENBL" button and dial in the known sensor resistance (389.8).
11______
Release the "ENBL" button.
12______
Wait for the automatic set point cal procedure to complete - about 10 seconds.
13______
Set the 5525 SENSOR Test Fixture Load switch to the "Chk" position. Verify the resistance reading is
200.2 KΩ plus or minus 0.5 KΩ.
5
Be sure you are properly grounded whenever you make contact with the rear connector.
6
Calibration can be terminated at any time by pressing the "OUTPUT" button. This will stop cal; no cal constants
will be stored to the EEPROM.
7
It doesn't matter which value resistor is used first.
7
.
3-12