Rack Mounting; Power-Up Sequence; Introduction To The Ldt-5525 Front Panel - ILX Lightwave LDT-5525 User Manual

Hide thumbs Also See for LDT-5525:
Table of Contents

Advertisement

2.2.2 Rack Mounting

The LDT-5525 Series Precision Temperature Controller may be rack mounted by installing a rack mount flange on
either side of the enclosure. All rack mount accessory kits contain detailed mounting instructions. Refer to Section
1.3 for applicable rack mount accessory part numbers.

2.3 Power-Up Sequence

With the LDT-5525 Series Precision Temperature Controller connected to an AC power source, pressing the POWER
switch will supply power to the instrument and start the power-up sequence.
During the power-up sequence, the following takes place. For about two seconds all indicators light up, and all of the
7-segment displays indicate "8". Then all lamps are turned off for two seconds. Then, the sensor switch position is
displayed for two seconds. After this, the unit is configured to the state it was in when the power was last shut off
(except for the display mode which defaults to I TE measurement). The adjust knob is always disabled at power up.

2.4 Introduction to the LDT-5525 Front Panel

The LDT-5525 Temperature Controller's front panel contains displays and controls for the Temperature Controller
hardware. Each of the labeled areas on the front panel (i.e. DISPLAY or MODE) is described in this chapter.
Refer to Figure 2.1 for the following discussions of the LDT-5525 Temperature Controller front panel sections. The
key words are in capital letters for quick identification.
2-2

Hide quick links: