H100B Pulse Oximeter User Manual
mode are set to 95 and 90 respectively; while they are
100 and 90 in Adu mode respectively.
Set the SpO
 Choose SpO
Function button to enter setup. The SpO
Alarm box will change from real line box to
broken line box. The adjustable range for
upper limit of SpO
limit of SpO
Alarm is set to less than 85, it will restore to
default value after the oximeter is turned on
again. In the NEO mode, if the value of SpO2
Hi Alarm is set to higher than 95, it will
restore to 95 after the oximeter is turned on
again.
 Press the Up Arrow or Down Arrow button to
increase or decrease values.
 Choose SpO
Function button to set it. The SpO
box will change from real line box to broken
line box. The adjustable range for the lower
limit of SpO
alarm limits as follows:
2
2
" to 100. If the value of SpO2 Hi
2
2
2
Hi Alarm in the menu, press the
is from "1 + the lower
2
Lo Alarm in the menu, press the
Alarm is from 0 to "the upper
- 45 -
Hi
2
Lo Alarm
2