Alarm Levels
In terms of severity, the oximeter's alarms levels can be
classified into two categories: high level alarms and
medium level alarms.
1. High level alarms
Indicate that the patient is in a life threatening
situation and an emergency treatment is demanded.
2. Medium level alarms
The patient's vital signs appear abnormal or the
oximeter system status appears abnormal, indicate
that prompt operator response is required.
The levels for both technical alarms and physiological
alarms are predefined and can not be changed by the
user.
Alarm Categories Table
Physiological
alarms
Technical alarms
High level alarms
SpO
Too High
2
SpO
Too Low
2
PR Too High
PR Too Low
- 53 -
Medium Level
Alarms
SpO
Sensor
2

Hide quick links: