AGC{sp}[1,2]{sp}<value>{cr} Set the gain in AGC OADM mode for each stage.
Example usage:
AGC{sp}1{sp}18{cr}
Reply is:
AGC{sp}1{sp}18{cr}
the gain for stage 1 in AGC OADM mode is set to 18 dB.
ALS{sp}[1,2]{sp} [E;D] {cr}
Set the automatic laser shutdown function to enable or disable when the loss of input alarm .
Set the output power value for each stage when the loss of input alarm or OPMI is activated.
Example usage:
ALS{sp}1{sp}E{cr}
Reply is:
ALS{sp}1{sp}E{cr}
set the automatic laser shutdown function for stage 1 to enable. If the input power goes below the
Loss of input value, the electronic will shutdown the laser diodes and then the output power will
be less than 0 dBm
Example usage:
ALS{sp}2{sp}D{cr}
Reply is:
ALS{sp}2{sp}D{cr}
set the automatic laser shutdown function for stage 2 to disable. If the input power stage 2 goes
below the Loss of input value stage 2, the output power stage 2 will stay at its previous value.
ALS {sp}[1,2]{cr}
Example usage:
ALS{sp}1{cr}
Reply is:
ALS{sp}1:{sp}ENABLE{cr}
Report the automatic laser shutdown function for stage 1 is enable.
Example usage:
ALS{sp}2{cr}
Reply is:
ALS{sp}2:{sp}DISABLE{cr}
EDFA optical amplifier – user manual
Set the automatic laser shutdown function.
Report the automatic laser shutdown function.