Parameter Type

The commands contains 5 kinds of parameters, different parameters has different
setting methods.
1. Boolean
The parameter should be "OFF", "ON", "0" or "1". For example:
:DISPlay:PERSist {{1|ON}|{0|OFF}}
"ON" and "1" denotes trun on (enable) the function, "OFF" and "0" denotes turn
off (disable) the fucntion.
2. Consecutive Integer
The parameter should be a consecutive integer. For example:
:DISPlay:BRIGhtness <ncount>
<ncount> could be the integer between 0 and 100 (including 0 and 100).
3. Consecutive Real Number
The parameters can be any value only in effective range precision permitting.
For example:
:TRIGger:SENSitivity <count>
<count> could be any value between 0.1 and 1 (including 0.1 and 1).
4. Discrete
The parameters can only be the cited value. For example:
:ACQuire:AVERages <count>
<count> could only be 2, 4, 8, 16, 32, 64, 128, 256.
5. ASCII Character String
The parameter should be composed of ASCII character string. For example:
:TRIGger:MODE <mod>
<mod> could be EDGE, PULSe, VIDEO, PATTern or ALTernation.
1-6
Programming Introduction
Programming Guide for DS1000B Series