Peak Detect - Rigol DS1204B Programming Manual

Ds1000b series digital oscilloscope
Hide thumbs Also See for DS1204B:
Table of Contents

Advertisement

RIGOL
For details about storage format of waveform points, please refer to Page 2-76:

Peak Detect

4.
:WAVeform:POINts:MODE
Command Format:
:WAVeform:POINts:MODE <points_mode>
:WAVeform:POINts:MODE?
Function:
This command sets the mode of waveform points. <points_mode> can be:
NORMal, MAXimum or RAW.
NOTE: What will be returned by :WAVeform:POINts? in different modes:
NORMal: Return data points currently display on the screen (600 points).
RAW: Return the data points of the memory data (in STOP state). In RUN state, no
data are returned, the system error code is 67 which indicate system condition is not
met and the execution failed.
MAXimum: Return the maximum valid data points in current state. In RUN state, the
screen data points are returned, while in STOP state, the memory data points are
returned.
MATH
FFT
CHx
Half-Channel
CHx
Returned Format:
The query returns NORMal, MAXimum or RAW.
Example:
:WAV:POIN:MODE NORM
:WAV:POIN:MODE?
2-72
NORmal
Normal
Peak
/Average
Detect
600
1200
500
500
600
1200
600
1200
Set the mode as NORMal.
Return NORMal.
RAW
MAX
In RUN state,
MAX is the same
with NORMal;
600
in STOP state,
500
MAX is the same
8192
with RAW.
16384
Programming Guide for DS1000B Series