Pulse Sr (1 Pulse Sync Reset) - Toshiba IK-TU61 Instruction Manual

Camera control unit
Table of Contents

Advertisement

( ( ( ( 2) ) ) ) 1 PULSE SR( ( ( ( 1 Pulse Sync Reset) ) ) )
Charge begins to accumulate after the trigger input operation to the DC IN/SYNC terminal or trigger signal
input (LVDS) to the DIGITAL terminal, the vertical sync signal is reset and field images are output. The
exposure time is set to 0.06 to 20 ms on the menu.
Trigger pulse level
(DC IN/SYNC terminal)
Trigger pulse fetch timing
Trigger pulse width
Trigger pulse interval
TRIGGER
(DC IN/SYNC terminal)
or
TRIGGER D
(DIGITAL terminal)
Internal VD
VD output
Index output
INDEX
Video output
: (Low Level):Less than 0.5V , (High Level):3.4V to 5V
: Rising edge / Falling edge selectable
: More than 2μs
: More than 42 ms
TRIGGER : Approx. 1.9μs
TRIGGER D : Approx. 2.4μs
Exposure period
Exposure end
16 to 17H
Approx. 4.4μs
n-1
n-1
Camera through image
- 44 -
Falling edge mode
Rising edge mode
Memory image