BK Medical 8805 User Manual

BK Medical 8805 User Manual

Linear array transducer

Advertisement

Quick Links

Linear Array Transducer
Type 8805
User Guide
English BB0337-F
August 2006

Advertisement

Table of Contents
loading

Summary of Contents for BK Medical 8805

  • Page 1 Linear Array Transducer Type 8805 User Guide English BB0337-F August 2006...
  • Page 2 WORLD HEADQUARTERS Mileparken 34 DK-2730 Herlev Tel.:+45 44528100 Fax:+45 44528199 www.bkmed.com Email: info@bkmed.dk B-K Medical Customer Satisfaction Input from our customers helps us improve our products and services. As part of our customer satisfaction program, we contact a sample of our customers a few months after they receive their orders.
  • Page 3: Table Of Contents

    Linear Array Transducer Type 8805 Introduction Scanning Plane General Information Caring for the Transducer Cleaning and Disinfection Starting Scanning Connecting the Transducer Changing Frequency Using a Transducer Cover Using the Transducer Control Button Changing Orientation Puncture Facilities UA1251 Performing Puncture and Biopsy...
  • Page 5: Introduction

    Linear Array Transducer Type 8805 Linear Array Transducer Type 8805 Introduction This is the user guide for Linear Array Transducer Type 8805 and must be used together with Transducer Care, Cleaning & Safety which contains important safety information. 8805 is suitable for musculoskeletal examinations.
  • Page 6: Caring For The Transducer

    8805 • Cleaning and Disinfection WARNING If at any time the scanner malfunctions, or the image is severely distorted or degraded, or you suspect in any way that the scanner is not functioning correctly: • Remove all transducers from contact with the patient.
  • Page 7: Connecting The Transducer

    8805 • Starting Scanning Connecting the Transducer WARNING Keep all plugs and sockets absolutely dry at all times. The transducer is connected to the scanner using the array Transducer Socket on the scanner. To connect, the transducer plug’s locking lever should first be in a horizontal position.
  • Page 8: Changing Orientation

    To change the orientation of the image on the monitor, refer to the applicable scanner user guide for instructions. Puncture Facilities Puncture and biopsy is possible with Type 8805. The puncture attachment is illustrated in the following pages with a brief description of its use and operating instructions. WARNING! It is essential for the patient’s safety that only the correct puncture attachment is used with...
  • Page 9 8805 • Puncture Facilities Fig. 3. Puncture attachment UA1251 Fig. 4. Type 8805 with puncture attachment UA1251 mounted...
  • Page 10: Performing Puncture And Biopsy

    8805 • Performing Puncture and Biopsy Performing Puncture and Biopsy WARNING It is essential for the patient’s safety that only the correct puncture attachments, as described in this guide, are used. Never use unauthorized combinations of transducers and puncture attachments or other manufacturers puncture attachments.
  • Page 11: Cleaning After Puncture And Biopsy

    8805 • Disposal Fig. 5. Puncture line for UA1251 with Type 8805 The transducer and puncture attachment can be removed without disturbing the position of the needle during puncture. Cleaning after Puncture and Biopsy If biological materials are allowed to dry on the transducer or puncture attachments, disinfection and sterilization processes may not be effective.