Section 2: Getting Started
Figure 2-24
Reinserting and securing the current input terminal fuse holder
WARNING
To change the 7A current input fuse on the rear panel:
Step 1: Disconnect the AC power
Verify that the meter is disconnected from AC power, as shown in
Figure 2-25
Disconnecting the AC power
Step 2: Release the current input terminal fuse holder
Using a flat-blade screwdriver, turn the 7A current input fuse holder on the rear panel to the left
until it is released (see
2-16
Before reconnecting the power line cord to your multimeter, make sure that
the power switch is in the "POWER OFF" position.
Figure
2-26).
Return to
Section Topics
Model 2100 6 1/2-Digit Resolution Digital Multimeter User's Manual
Figure
2-25.
2100-900-01 Rev. D / September 2011