Model 2100 6 1/2-Digit Resolution Digital Multimeter User's Manual
[SENSe:]TEMPerature:RTD:RZERo? [MINimum|MAXimum]
Query the R-Zero constant for the user-defined RTD type.
[SENSe:]TEMPerature:RTD:ALPHa {<value>|MINimum|MAXimum}
Set the alpha constant for the user-defined RTD type.
[SENSe:]TEMPerature:RTD:ALPHa? [MINimum|MAXimum]
Query the alpha constant for the user-defined RTD type.
[SENSe:]TEMPerature:RTD:BETA {<value>|MINimum|MAXimum}
Set the beta constant for the user-defined RTD type.
[SENSe:]TEMPerature:RTD:BETA? [MINimum|MAXimum]
Query the beta constant for the user-defined RTD type.
[SENSe:]TEMPerature:RTD:DELTa {<value>|MINimum|MAXimum}
Set the delta constant for the user-defined RTD type.
[SENSe:]TEMPerature:RTD:DELTa? [MINimum|MAXimum]
Query the delta constant for the user-defined RTD type.
[SENSe:]TEMPerature:SPRTD:RZERo {<value>|MINimum|MAXimum}
Set the sensor R value at 0 degrees Celsius.
[SENSe:]TEMPerature:SPRTD:RZERo? [MINimum|MAXimum]
Query the sensor R value at 0 degrees Celsius.
[SENSe:]TEMPerature:SPRTD:A4 {<value>|MINimum|MAXimum}
Set the A4 coefficient.