[SENSe:]<function>:RANGe? [MINimum|MAXimum]
Query the range for the selected function. For frequency and period, use FREQuency:VOLTage
or PERiod:VOLTage. Autorange thresholds: Down range at <10% of range; Up range at >120% of
range.
[SENSe:]<function>:RANGe:AUTO {OFF|ON}
Disable or enable autorange for the selected function.
[SENSe:]<function>:RANGe:AUTO?
Query the autorange setting. Return "1" (ON) or "0" (OFF).
[SENSe:]<function>:RESolution {<resolution>|MINimum|MAXimum}
Set resolution for the selected function (not valid for frequency, period, or ratio). Specify the
resolution in the same units as the measurement function, not in number of digits. MIN selects the
smallest value accepted, which gives the most resolution. MAX selects the largest value accepted,
which gives the least resolution.
[SENSe:]<function>:RESolution? [MINimum|MAXimum]
Query the resolution for the selected function.
[SENSe:]UNIT {Cel|Far|K}
Select units for temperature measurement. Cel means Celsius, Far means Fahrenheit, and K
means Kelvin.
[SENSe:]UNIT?
Query units for temperature measurement.
[SENSe:]TEMPerature:RTD:TYPE{PT100|D100|F100|PT385|PT3916|USER|SPRTD}
Select the RTD type for temperature measurement with RTD.
[SENSe:]TEMPerature:RTD:TYPE?
Query the RTD type for temperature measurement with RTD.
[SENSe:]TEMPerature:RTD:RZERo {<value>|MINimum|MAXimum}
Set the R-Zero constant for the user-defined RTD type.