Model 2100 6 1/2-Digit Resolution Digital Multimeter User's Manual
CALCulate:STATe?
Query the state of the math function. Returns "0" (OFF) or "1" (ON).
CALCulate:PERCent:TARGet {<value>|MINimum|MAXimum}
Set the target value for percent math function. The multimeter clears the value when Min/Max is
turned on, when the power has been off or a remote interface reset.
CALCulate:PERCent:TARGet? [MINimum|MAXimum]
Query the target value for the PERCent math function.
CALCulate:AVERage:MINimum?
Read the minimum value found during the Min/Max operation. The multimeter clears the value
when Min/Max is turned on, when the power has been turned-off, or a remote interface has been
reset.
CALCulate:AVERage:MAXimum?
Read the maximum value found during the Min/Max operation. The multimeter clears the value
when Min/Max is turned on, when the power has been turned-off, or a remote interface has reset.
CALCulate:AVERage:AVERage?
Read the average value of all readings taken since the Min/Max operation was enabled. The
multimeter clears the value when Min/Max is turned on, when the power has been turned-off or a
remote interface has been reset.
CALCulate:AVERage:COUNt?
Read the number of readings taken since Min/Max was enabled. The multimeter clears the value
when Min/Max is turned on, when the power has been turned-off ,or a remote interface has been
reset.
CALCulate:NULL:OFFSet {<value>|MINimum|MAXimum}
Store a null value in the multimeter's null register. You must turn on the math operation before
writing to the math register. You can set the null value to any number between 0 and 120% of the
highest range.
CALCulate:NULL:OFFSet?
Query the null value.