[SENSe:]TEMPerature:SPRTD:DX? [MINimum|MAXimum]
Query the D coefficient.
[SENSe:]<function>:NPLCycles {0.02|0. 1|1|10|MINimum|MAXimum}
Set the integration time in number of power line cycles for the selected function. This command is
valid only for DCV, DCI, 2-wire ohms, and 4-wire ohms.
[SENSe:]<function>:NPLCycles? [MINimum|MAXimum]
Query the integration time for the selected function.
[SENSe:]FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
Set the gate time (or aperture time) for frequency function. Specify 10ms (4 1/2 digits), 100ms
(default; 5 1/2 digits), or 1 second (6 1/2 digits).
[SENSe:]FREQuency:APERture? [MINimum|MAXimum]
Query the gate time (or aperture time) for frequency function.
[SENSe:]PERiod:APERture{0.01|0.1|1|MINimum|MAXimum}
Set the gate time (or aperture time) for period function. Specify 10ms (4 1/2 digits), 100ms (default;
5 1/2 digits), or 1 second (6 1/2 digits).
[SENSe:]PERiod:APERture? [MINimum|MAXimum]
Query the gate time (or aperture time) for period function.
[SENSe:]DETector:BANDwidth {3|20|200|MINimum|MAXimum}
Specify the lowest frequency expected in the input signal. The meter selects the slow, medium, or
fast AC filter based on the frequency you specify.
[SENSe:]DETector:BANDwidth? [MINimum|MAXimum]
Query the AC filter and return the bandwidth.
[SENSe:]ZERO:AUTO {OFF|ONCE|ON}
Disable or enable the AUTO ZERO mode. The OFF and ONCE modes have a similar effect. OFF
mode does not issue a new offset measurement until the multimeter goes to the "WAIT-FOR-
TRIGGER" state. Parameter ONCE issues an immediate offset measurement.