Temperature too low / too high: This message ap-
pears at the start of a test if the ambient temperature
or the temperature of the meter is between +8 and
+10 °C or between +40 and +42 °C.
use
, to select whether you wish to Test or Quit.

Test:
in this case, the meter advances a test area.

Quit:
in this case, the meter returns to the Main
menu. Close the tip cover.
If you decide to perform the test in spite of the warn-
ing, the result will be flagged and saved with the sym-
bol .
do not use blood glucose results obtained follow-
ing these warnings as a basis for making thera-
peutic decisions. These results may be incorrect.
Incorrect results can cause the wrong therapy
recommendation to be made and thus produce
serious adverse health effects. Move to a place
where the temperature is between +10 and
+40 °C and wait for the temperature of the me-
ter to adjust to this temperature.