15
Testing and storage conditions
do not use blood glucose results obtained
following these warnings as a basis for mak-
ing therapeutic decisions. These results may
be incorrect. Incorrect results can cause the
wrong therapy recommendation to be made
and thus produce serious adverse health
effects.

Tests cannot be performed at temperatures below
+8 °C and above +42 °C. In this event, the following
error message appears:
Never try to speed up the cooling or warming
of your meter, e.g. by placing it in the refriger-
ator or on a radiator. doing so can damage
the meter and cause it to deliver incorrect
results. Incorrect results can cause the wrong
therapy recommendation to be made and
thus produce serious adverse health effects.
218

do not store the lancing device in a primed state and
do not store it for any length of time in a place
where the temperature is very low or high, e.g. in a
hot car. This can cause parts of the lancing device to
become warped, so impairing its operation.