did you perform the control test according to
the user's Manual?
did you wait for the application prompt to be
displayed before applying control solution?
did you apply the control solution immedi-
ately after opening the applicator?
did you enter the correct color?
did you perform the test within the correct
temperature range (10–40 °C)?
Has the expiry date for the control solution
yes
been reached (see label on the bottom of the
vial next to the symbol )?
yes
did you read the section "Storage and stabil-
ity of control solutions" in the package insert
yes
for the control solution?
did you observe the proper storage condi-
yes
tions for the meter, test cassette and control
solutions (see Chapter 15 and the test cas-
yes
sette and control solution package inserts)?
If your meter has been dropped, that can also lead to
control test results which are not within the concentra-
tion range displayed and for which Control test not OK
is displayed. In this case, you should also call
Accu-Chek Customer Care.