Monitor Settings for White Balance

Monitor Settings The monitor settings are decisive for the image you see on the monitor. Set your
Note
Note

Executing White Balance

Specialized
microscopy
techniques
Transmitted Light When using the transmitted light method, look for a position through which the
Reflected light When using the reflected light method, you need a reference surface.
Related Topics
White Balance 17
monitor to the appropriate color temperature, should your device allow this. To
do this, compare the image's color impression on the monitor and in the
eyepiece.
The color temperature setting can be altered on most monitors via the monitor
menu buttons located on the monitor. In other cases, the color temperature can
be altered via the operating system.
Changing the color temperature using the operating system
1)
Right click on Desktop.
The Desktop's context menu is opened.
2)
Select the entry
Properties
The
Display Properties
3)
Select the
Settings
4)
Click the
Advanced
Another
Display Properties
function of this dialog box depends on the graphics card installed on
your system and your monitor.
Search for a tab with the name
The appearance of the operating elements might differ from the one being
described here because the adaptation of the color temperature is not part of the
operating system, but rather is made accessible by the monitor manufacturer or
the graphics card being used.
5)
Select a color temperature whose color impression best corresponds to that
seen in the eyepiece.
Should your system offer a slide control with the options
do not alter your color temperature with these.
It may be difficult to execute a white balance in some microscopic methods, e.g.
DIC or polarization. In these cases, execute the white balance using the bright
field method and only then change to the special methods.
light passes through the cover slip, embedding material, and microscope slide,
but not through the specimen. Should such a position not exist, remove the
sample and carry out the white balance without a sample.
Standardized gray cards are best for this. Should one not be available, you can
also use white laboratory porcelain or neutral ND-filter paper. Should your
samples contain very reflective areas, defocus your microscope when setting the
white balance.
from the context menu.
dialog box is opened.
tab.
button.
dialog box is opened. The appearance and
Color
or
Camera and light microscope
Color
Management.
warm
and cold,