BC Biomedical PS-2200 series User Manual page 8

Multi-parameter patient simulator
Table of Contents

Advertisement

PS-2200
ECG-PERFORMANCE FUNCTIONS
The unit will generate Sine, Square, Triangular and Pulse waveforms with adjustable
amplitudes for performance testing. A special Automatic mode is available to auto
sequence through the entire range of waveforms. Also, a R-Wave (Havertriangle)
waveform may be selected with separate amplitude and width settings.
• SINE:
• SQUARE:
• TRIANGLE:
• PULSE:
• AMPLITUDE:
• AUTOMATIC MODE
• R WAVE RATE:
• R WAVE WIDTH:
• R WAVE AMPLITUDE:
RESPIRATION
Respiration is simulated at 8 different rates, with the ability to select from 4 Baseline
Impedances, the Lead in which it will appear, and the Delta Ohms (amplitude) of the signal.
Additionally, an Apnea period may be selected.
• RATE:
• BASELINE IMPEDANCE:
• LEAD:
• DELTA IMPEDANCE:
• APNEA:
0.1, 0.5, 5, 10, 40, 50, 60,100 Hz
0.125, 2 Hz
2, 2.5 Hz
30,60,120 BPM; 60 ms WIDTH
.05,.10,.15,.20,.25,.30,.35,.40,.45,.50,
1.0,1.5,2.0,2.5,3.0,3.5,4.0,4.5,5.0 mV (Lead II)
30,60,80,120,200,250 BPM
8,10,12,20,30,40,50,60,70,80,90,100,110,120,130,
140,150,160,170,180,190,200 ms
.05,.10,.15,.20,.25,.30,.35,.40,.45,.50,1.0,1.5,2.0,2.5,3.0,3.5,
4.0,4.5,5.0 mV (Lead II)
15,20,30,40,60,80,100,120 BrPM
500,1000,1500,2000 OHMS
LA or LL
0.1,0.2,0.5,1.0,2.0,3.0 OHMS
12, 22 and 32 SECONDS PLUS CONTINUOUS
6