BC Biomedical PS-2200 series User Manual page 34

Multi-parameter patient simulator
Table of Contents

Advertisement

PS-2200
The pacemaker rhythms and signals can be selected by using
the parameter to change and using
is used to accept the new setting.
Alternately, to see a submenu of all the options for a highlighted parameter, use
Use
to scroll to the desired option. Then
setting.
The following is a brief description of how the PS-2200 simulates the available Pacemaker
Waveforms:
Abbreviation
Atrial Pacer
Asynchronous
Non-Capture
Non-Function
Demand –
Occasional
Demand –
Frequent
AV – Sequential
PACEMAKER
Waveform
Atrial Pacemaker Wave
Asynchronous Pacemaker Wave
Ventricular Pacemaker Wave
with Periodic Non-Response
Ventricular Pacemaker Wave
with no Heart Response
Demand Pacemaker Wave with
Occasional Sinus Beats
Demand Pacemaker Wave with
Frequent Sinus Beats
AV-Sequential Pacemaker Wave
to scroll to the desired option. Then
80 BPM with Pacer Pulse at the start
of each P wave
75 BPM with Pacer Pulse at the start
of each QRS wave and no P wave
75 BPM Ventricular Paced beats
with every tenth beat not responding
75 BPM Ventricular Paced beats
with no heart response
20 NSR beats followed by 20
Ventricular Paced beats
40 NSR beats followed by 40
Ventricular Paced beats
75 BPM with Pacer Pulse at the start
of both the P and QRS waves
32
to highlight
is used to accept the new
Description
.