BC Biomedical PS-2200 series User Manual page 74

Multi-parameter patient simulator
Table of Contents

Advertisement

PS-2200
Setting Up a Test
1) The Device Under Test (DUT, Cardiac Output Monitor) needs to be setup to match the
PS-2200. The following settings are required:
Catheter
Calibration Coefficient
Injectate Volume
Injectate Temperature
2) Connect Com2100 Module to the CO/Temp input on the PS-2200.
3) Connect the Blood Temperature Sensor Line (BT Thermistor Cable) from the DUT to
the small 4 pin connector on the COM2100.
4) Connect the Injectate Temperature Sensor Line for the DUT to the larger 4 pin
connector on the MCO-2100.
5) Turn on the DUT and the PS-2200. The DUT should show a blood temperature of
about 37 °C.
6) Adjust the trim pot on the MCO-2100 Module until the DUT shows the desired injectate
temperature (0 °C or 24 °C)
Baxter Edwards 93a – 131 – 7f
0.542 for 0 °C Injectate
0.595 for 24 °C Injectate
10 cc
0 °C or 24 °C
72