BC Biomedical NIBP-1010 User Manual

BC Biomedical NIBP-1010 User Manual

Non-invasive blood pressure simulators

Advertisement

NON-INVASIVE

BLOOD PRESSURE

SIMULATORS
NIBP-1030
NIBP-1020
NIBP-1010
USER MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for BC Biomedical NIBP-1010

  • Page 1: Blood Pressure

    NON-INVASIVE BLOOD PRESSURE SIMULATORS NIBP-1030 NIBP-1020 NIBP-1010 USER MANUAL...
  • Page 3: Table Of Contents

    BC BIOMEDICAL NIBP-1000 SERIES TABLE OF CONTENTS WARNINGS, CAUTIONS, NOTICES ................ii DESCRIPTION......................1 OVERVIEW........................4 KEYS..........................7 SCREENS........................8 SETUP ......................... 13 OPERATIONS......................16 CONNECTING PRESSURE CUFFS............... 16 CONNECTING PATIENT LEADS ..............17 IBP CONNECTOR (NIBP-1030 ONLY)............17 TEMPERATURE CONNECTOR (NIBP-1030 ONLY)........18 AUX CONNECTOR..................
  • Page 4 WARNING - USERS The NIBP-1000 Series Simulators are for use by skilled technical personnel only. WARNING - USE The NIBP-1000 Series Simulators are intended for testing only and they should never be used in diagnostics, treatment or any other capacity where they would come in contact with a patient.
  • Page 5 CAUTION - SERVICE The NIBP-1000 Series Simulators are intended to be serviced only by authorized service personnel. Troubleshooting and service procedures should only be performed by qualified technical personnel. CAUTION - ENVIRONMENT The NIBP-1000 Series Simulators are intended to function between 15 and 40 °C. Exposure to temperatures outside this range can adversely affect the performance of the Simulators.
  • Page 6 NOTICE – CE The NIBP-1000 Simulators bear the mark Based on the following testing standards: ELECTROMAGNETIC COMPATIBILITY DIRECTIVE EMC – Directive 89/336/EEC and 2004/108/EC as amended by 92/31/EEC, 93/68/EEC and Directive 91/263/EEC [ TTE/SES ] EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003 “Electrical equipment for measurement, control and laboratory use –...
  • Page 7 NOTICE – SYMBOLS Symbol Description Caution (Consult Manual for Further Information) Center Negative Direct Current Per European Council Directive 2002/95/EC, do not dispose of this product as unsorted municipal waste. NOTICE – ABBREVIATIONS Alternating Current Blood Pressure Beats per minute brpm breaths per minute Celsius...
  • Page 8 POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES. NOTICE – CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL, INC. PO BOX 25125 9415 GENTRY AVE ST. LOUIS, MO 63125...
  • Page 9: Description

    BP waveform. The following are highlights of some of the main features: NIBP-1010 (BASIC FEATURES, NIBP ONLY): • LARGE GRAPHICS DISPLAY WITH CURSOR SELECTION OF OPTIONS AND SETUP OF PARAMETERS •...
  • Page 10 NIBP-1030 (NIBP, IBP, TEMPERATURE, ECG PACE & RESP) HAS ALL THE 1020 MODEL FEATURES PLUS: • BACKLIT LCD DISPLAY • SYNCHRONIZED INVASIVE BLOOD PRESSURE OUTPUT • SELECTABLE IBP SENSITIVITY 5/40 µV/V/mmHg • STATIC IBP SIMULATION –10mmHg TO 400 mmHg • LEAK RATE TEST •...
  • Page 11 INVASIVE BLOOD PRESSURE CABLES FOR A COMPLETE LIST CONSULT THE BC GROUP WEBSITE AT WWW.BCGROUPINTL.COM TEMPERATURE CABLES BC20 - 41333 UT-1 YSI-400 BC20 - 41334 UT-2 YSI-700...
  • Page 12: Overview

    OVERVIEW This section looks at the layout of the NIBP-1030 and gives descriptions of the elements that are present. 10 Snap Patient Lead Connectors: Male Quick Connect for Pressure LCD Graphical Input Display: Shows Test Data and Configuration Data 6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit...
  • Page 13 This section looks at the layout of the NIBP-1020 and gives descriptions of the elements that are present. 10 Snap Patient Lead Connectors: Male Quick Connect for Pressure LCD Graphical Input Display: Shows Test Data and Configuration Data 6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit...
  • Page 14 This section looks at the layout of the NIBP-1010 and gives descriptions of the elements that are present. Male Quick LCD Graphical Connect for Display: Pressure Shows Test Data and Input Configuration Data 6 Light Touch Keys for Selecting Parameters...
  • Page 15: Keys

    KEYS Six tactile-touch keys are provided for system operation: – This key turns the unit off and on. The unit will return to the screen that was active when it was turned off. – In the DISPLAY MODE, these keys toggle the display through the available main screens.
  • Page 16: Screens

    SCREENS MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional). The available screens can be toggled using PRESSURE ONLY SCREEN – This screen has a large pressure display, as shown below. Also displayed on this screen is the peak pressure and selected output waveform The display will resemble the following: PEAK PRESSURE...
  • Page 17 OUTPUT WAVEFORM SCREEN – This screen shows the output waveform, as well as, the pressure, peak pressure and output selection. NOTE This waveform is not intended to be physiologically correct. The display will resemble the following: PRESSURE PEAK PRESSURE (Optional) OUTPUT WAVEFORM (as generated by...
  • Page 18 PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as, the pressure, peak pressure and output selection PRESSURE PEAK PRESSURE (Optional) PRESSURE GRAPH 2mmHg per pixel OUTPUT SELECTION Respiration Systolic/Diastolic (mean) Mean Heart Rate Rate (Optional) Adult 120/80 Adult 120/80 w/Pace Pacemaker 70...
  • Page 19 ECG OUTPUT SCREEN – This screen shows the selected ECG output mode. (NIBP-1020 & NIBP-1030 units only) NOTE: While in this mode, the NIBP simulation does not run. NOTE: NSR ECG output is active during NIBP testing at the rate stipulated for the selected test.
  • Page 20 Batteries charge from provided AC power adapter. The charge time is about 4 hours from full discharge. While the batteries are charging, the display will flash “charging.” CAUTION USE ONLY THE AC POWER ADAPTER PROVIDED WITH THE NIBP-1010, NIBP-1020 AND NIBP-1030.
  • Page 21: Setup

    SETUP The Setup Mode allows the user to adjust the configuration of the simulator. The Setup screen can be entered using the key. The parameters can be changed by using key to highlight the line and to toggle the available options. The Setup screen can be exited using the key.
  • Page 22 The following is a breakdown of the parameters available in the configuration of the unit and their available options: System Setup Configuration Parameter Description Range Adjusts the Systolic Output of the NIBP Systolic Shift ±50.0 Simulation. This is not a direct mmHg adjustment. Adjusts the Diastolic Output of the NIBP Diastolic Shift ±50.0...
  • Page 23 SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option to shift test results to compensate for different methods of measuring Oscillometric NIBP by various manufacturers and models of devices under test. CAUTION These adjustments must be used with caution as they will allow the user to adjust the output results to invalid values.
  • Page 24: Operations

    OPERATIONS CONNECTING PRESSURE The NIBP-1010, NIBP-1020 and NIBP-1030 Blood Pressure Simulators are connected between the blood pressure cuff and the monitor (Device Under Test, DUT). The blood pressure cuff should be disconnected from the DUT and a ‘T’ adapter inserted between the cuff and the DUT.
  • Page 25: Connecting Patient Leads

    CONNECTING PATIENT LEADS For the models with ECG output, test snaps are provided along the sides and are identified by the markings on the overlay. IBP CONNECTOR (NIBP-1030 ONLY) For the models with IBP output, connect a standard BP test cable (mini-din to monitor connector) between the NIBP-1030 and the DUT (Device Under Test).
  • Page 26: Temperature Connector (Nibp-1030 Only)

    TEMPERATURE CONNECTOR (NIBP-1030 ONLY) For the models with Temperature output, connect a standard Temperature test cable (mini- din to monitor connector) between the NIBP-1030 and the DUT (Device Under Test). The pin-out is shown below. AUX CONNECTOR The Aux connector is used for both serial communications and SpO The serial connection is made using the BC20-41337 cable.
  • Page 27: Theory Of Operations

    THEORY OF OPERATIONS NIBP (Non-Invasive Blood Pressure) Function A digital stepping motor and piston assembly is used to generate the output waveforms. A differential pressure sensor is used to measure the pressure of the cuff. The pressure sensor is read by a 16-bit differential Analog to Digital converter. The cuff pressure is monitored and the motor/piston assembly reacts to provide the correct response based on the selected mode.
  • Page 28 ECG Function The ECG output is performed by a 12-bit Digital to Analog converter. All waveforms for the standard ECG and Arrhythmias are generated via data stored in the program. Respiration Function A 12-bit Digital to Analog converter is used to generate the respiration waveform. The respiration output is present on the LA ECG lead only.
  • Page 29: Running A Test

    RUNNING A TEST Depending on the model, there are up to 9 Basic Test Modes and 3 Sub-Test Modes. This section will walk through each of the tests and their basic operation. The Main Tests are accessible with a single key. The key will scroll through the following tests in a continuous loop: Adult 120/80...
  • Page 30 There are 5 selectable NIBP Basic Test Modes; Adult, Adult w/pace, Adult High, Adult Low and Neonatal. The displays will resemble the following examples: NOTE: The screens for the first test will have the individual components labeled. The component labels for subsequent tests are the same. Adult 120/80 (Typical Screens) PEAK PRESSURE...
  • Page 31 Adult 120/80 with Pace (Typical Screens) Adult High 120/80 (Typical Screens)
  • Page 32 Adult Low 80/40 (Typical Screens) Neonatal 70/40 (Typical Screens)
  • Page 33 ALARM TEST: Use the key to cycle through the Test Modes until Alarm Test appears. The purpose of this function is to help test the alarms of the monitor under test. The ECG output will alternate from a 47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR with 80 brpm respiration.
  • Page 34 ARRHYTHMIA SEQUENCE: (NIBP-1030 Only) Use the key to cycle through the Test Modes until Arrhythmia Sequence appears. The purpose of this function is to provide a timed sequence of some of the more common Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed below.
  • Page 35 LEAK TEST: (NIBP-1030 Only) Use the key to cycle through the Test Modes until Leak Test appears. The purpose of this function is to provide a standard pressure leak test. The pressure input is connected to the system that is to be monitored. Pressure is then applied to the system. The Leak Test is initiated by pressing the key.
  • Page 36 MANOMETER: Use the key to cycle through the Test Modes until Manometer appears. The purpose of this function is to allow the unit to be used strictly as a Manometer with a full scale range of +/- 500 mmHg. NOTE: The Graphic Screens are also available in this mode for maximum flexibility The displays will resemble the following examples: (Typical Screens) PEAK PRESSURE PRESSURE...
  • Page 37 SUB TEST MODES PEAK: (Optional) The optional Peak test feature is available whenever the display is in a pressure mode. It will continuously monitor the pressure input and display and hold the maximum value. This value may be reset at any time by pressing the key.
  • Page 38 ECG PERFORMANCE TEST: (NIBP-1020 & NIBP-1030) The ECG output is selected by pressing until the ECG Waveforms menu is displayed. The key is used to choose the desired ECG output. There are 16 fixed waveforms available. NOTE: While in this mode, the NIBP simulation is disabled, but the IBP simulation is active (NIBP-1030 only).
  • Page 39 DYNAMIC IBP: (NIBP-1030 Only) The IBP output matches the NIBP setting in the NIBP mode or the ECG setting in the ECG Waveform mode. NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in the SETUP Mode before using IBP. (See Setup section for more information.) STATIC IBP: (NIBP-1030 Only) A static Blood Pressure output may be selected in the SETUP mode.
  • Page 40 TEMPERATURE SIMULATION: (NIBP-1030 Only) The temperature setting can be selected in the SETUP mode. This fixed value will remain on the Temperature output until changed. The output will simulate both YSI 400 and YSI 700 Temperature probes. NOTE: Both outputs are available at the output connector simultaneously.
  • Page 41: Manual Revisions

    MANUAL REVISIONS Revision # Program # Revisions Made Rev 01 DT7355CA Origination Rev 02 DT7355CA Specification Information Updated Rev 03 DT7355CA Miscellaneous Edits Rev 04 DT7355CF Add Systolic/Diastolic adjustments Rev 05 DT7355CG Output Selection edited Rev 06 DT7355CI Miscellaneous Edits Rev 07 DT7355CI Warnings, Cautions, Notices Updated...
  • Page 42: Specifications

    SPECIFICATIONS NIBP-1010 Blood Pressure RANGE 0-500 mmHg @ 20°C ACCURACY +/- (1% of Reading + .5 mmHg) RATE 80, 94 bpm ACCURACY +/- 1% Physical DISPLAYS LCD Graphical 128 X 64 Pixels 7 x 5 x 4 Inches ENCLOSURE (177.8 x 127.0 x 101.6 mm) ABS Plastic <...
  • Page 43 NIBP-1020 Blood Pressure RANGE 0-500 mmHg @ 20°C ACCURACY +/- (1% of Reading + .5 mmHg) RATE 80, 94 bpm (synchronized to ECG) ACCURACY +/- 1% ECG NSR RATE 30,60,120,240 BPM ACCURACY +/- 1% AMPLITUDE 2.75 mV ACCURACY +/- 2% @ Lead II ECG Performance SINE WAVE 10,60,100 Hz...
  • Page 44 Respiration RATE ACCURACY +/- 1% IMPEDANCE DELTA 3.0 ohms ACCURACY +/- 10% BASELINE 1000 ohms ACCURACY +/- 5% Physical LCD Graphical 128 X 64 Pixels DISPLAYS White LED Backlight 7 x 5 x 4 Inches ENCLOSURE (177.8 x 127.0 x 101.6 mm) ABS Plastic <...
  • Page 45: Invasive Blood Pressure

    NIBP-1030 Blood Pressure RANGE +/- 500 mmHg @ 20°C ACCURACY +/- (1% of Reading + .5 mmHg) RATE 80, 94 bpm (synched to ECG) ACCURACY +/- 1% Invasive Blood Pressure -10,-5,0,20,40,50,60,80,100,120,150, STATIC PRESSURE 160,200,240,250,300,320,400 mmHg +/- (1% Full Range + 1 mmHg) or ACCURACY +/- (2% Setting + 2 mmHg) whichever is better...
  • Page 46 ECG Performance SINE WAVE 10,60,100 Hz SQUARE WAVE 0.125, 2.000 Hz TRIANGLE WAVE 2.000 Hz RATE ACCURACY +/- 1% AMPLITUDE 2.75 mV AMPLITUDE +/- 2% @ Lead II ACCURACY Pacemaker Waveforms AMPLITUDE 3 mV ACCURACY +/- 10% WIDTH 6 ms ACCURACY +/- 5% Respiration...
  • Page 47 Physical DISPLAYS LCD Graphical 128 X 64 Pixels 7 x 5 x 4 Inches ENCLOSURE (177.8 x 127.0 x 101.6 mm) ABS Plastic < 3 Lbs WEIGHT (< 1.36 Kg) FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE -20 to 65 C Electrical...
  • Page 48: Notes

    NOTES...

This manual is also suitable for:

Nibp-1030Nibp-1020

Table of Contents