:DIGital:DATA:INPut:NOISe:BANDwidth

Supported
N5172B/82B with option 004
:DIGital:DATA:INPut:NOISe:BANDwidth <value><unit>
:DIGital:DATA:INPut:NOISe:BANDwidth?
This command selects the flat noise bandwidth value of the real–time noise for an ARB
waveform. Typically, this value is set slightly wider than the signal bandwidth. The minimum
increment value is 0.001 Hz.
Example
:DIG:DATA:INP:NOIS:BAND 8MHZ
The preceding example sets the flat noise bandwidth to 8 MHz.
*RST
+1.00000000E+000
Range
1 Hz to 160 MHz (depends on the installed baseband generator option)
Flat Noise Bandwidth
Key Entry

:DIGital:DATA:INPut:NOISe:CBRate

Supported
N5172B/82B with option 004
:DIGital:DATA:INPut:NOISe:CBRate <1bps - 999Mbps>
:DIGital:DATA:INPut:NOISe:CBRate?
This command sets a value of the carrier bit rate (gross bit rate) for purposes of calculating the
E
/N
(energy per bit over noise power density at the receiver). When the carrier to noise ratio
b
0
format is set to E
/N
b
0
of the carrier bit rate will have an immediate impact on the carrier to noise ratio as specified by
E
/N
. The carrier bit rate is derived from the symbol rate and bits per symbol of the modulation.
b
0
The carrier bit rate is a saved instrument state that is recorded in the waveform header.
The query returns the current carrier bit rate setting.
Example
:DIG:DATA:INP:NOIS:CBR 5
The preceding example sets the carrier bit rate to 5 bps.
Default
1.000 bps
Range
1 bps to 999 Mbps
Carrier Bit Rate
Key Entry
Agilent X-Series Signal Generators SCPI Command Reference
(refer to the
:DIGital:DATA:INPut:NOISe:CNFormat
Digital Signal Interface Module Commands
Digital Subsystem—Option 003 and 004 ([:SOURce])
command), the adjustment
355

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents