*SAV

Supported
All Models
*SAV <reg>,<seq>
The Save (SAV) command saves signal generator settings to the specified memory register <reg> of
the specified sequence <seq>.
Range
Registers: 0 to 99
Save Reg
Key Entry
Remarks
The save function does not save all signal generator settings. Refer to the User's
Guide for more information on the save function. Refer to
information on recalling saved signal generator settings.

*SRE

Supported
All Models
*SRE <data>
The Service Request Enable (SRE) command sets the value of the Service Request Enable Register.
The variable <data> is the decimal sum of the bits that will be enabled. Bit 6 (value 64) is ignored
and cannot be set by this command.
Range
0 to 255
Remarks
Refer to the Programming Guide for more information.
Entering values from 64 to 127 is equivalent to entering values from 0 to 63.
The setting enabled by this command is not affected by signal generator preset or
*RST. However, cycling the signal generator power will reset it to zero.
*SRE?
Supported
All Models
*SRE?
The Service Request Enable (SRE) query returns the value of the Service Request Enable Register.
Range
0 to 63 or 128 to 191
Remarks
Refer to the Programming Guide for more information.
Agilent X-Series Signal Generators SCPI Command Reference
Sequences: 0 to 9
Save Seq[n] Reg[nn]
System Commands
IEEE 488.2 Common Commands
"*RCL" on page 114
for
115