The Data Display Setup screen includes check boxes,
radio buttons and drop-down menu fields. The Sample
screen changes to agree with the display setup.
Parameters such as font size can be different for different
configurations.

Setting the Method of Counting Data

The choices for data counting are:
Cumulative
Cumulative counting mode includes all particles that are
greater than or equal to the particle size selected in the
Sample Volume field.
Differential
Differential counting mode includes all particles that are
greater than or equal to the particle size selected in the
Sample Volume field, but less than the next largest
particle size.

Location Setup Submenu

Sample location can be recorded for the samples. If the
Product is used in multiple areas, it will make the samples
clear if the sample location was recorded. 75 locations
can be added.
To update or change a location parameter:
1.
From the primary Sample Setup Menu icon, push 
three times to show the Location Setup icon.
14
2.
Push  to enter the submenu.
3.
Use the Product buttons to choose an action for the
location setup. See Table 9.
4.
After the location data is changed, push  to return
to the navigation menu.
Table 9. Location Submenu Icons
Icon
Add a location
Edit a location
Delete a location
In the locations, use this to move a
selected location down in the list.
In the locations, use this to move a
selected location up in the list.
Description