Segmented Memory Re-Arm Time; Saving Data From Segmented Memory - Agilent Technologies InfiniiVision 4000 X Series User Manual

Table of Contents

Advertisement

12
Acquisition Control
The Analyze Segments softkey appears when the acquisition is stopped and
the segmented memory feature is on or when the serial Lister is enabled.
You can also turn on infinite persistence (in the Display Menu) and press
the Analyze Segments softkey to create an infinite persistence display.

Segmented Memory Re-Arm Time

After each segment fills, the oscilloscope re- arms and is ready to trigger
in about 1 µs.
Remember though, for example: if the horizontal time per division control
is set to 5 µs/div, and the Time Reference is set to Center, it will take at
least 50 µs to fill all ten divisions and re- arm. (That is 25 µs to capture
pre- trigger data and 25 µs to capture post- trigger data.)

Saving Data from Segmented Memory

You can save either the currently displayed segment (Save Segment -
Current), or all segments (Save Segment - All) in the following data formats:
CSV, ASCII XY, and BIN.
Be sure to set the Length control to capture enough points to accurately
represent the captured data. When the oscilloscope is busy saving multiple
segments, progress is displayed in the upper right area of the display.
For more information, see
page 312.
230
"To save CSV, ASCII XY, or BIN data files"
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide