A&D AD-4212A-100 Instruction Manual page 88

Production weighing unit
Table of Contents

Advertisement

Mass of the special weighing pan (AD-4212A series balance)
Design the weighing pan so that the mass falls in the ranges shown in the table below:
Model
AD-4212A-100
AD-4212A-200
AD-4212A-600
AD-4212A-1000
AD-4212A-100/200
The weighing range for each pan design is shown below:
AD-4212A-100
When the weighing pan
is removed
60 g
Mass range
to be weighed
110 g
70 g
Mass
of special
20 g
weighing pan
Pan support is used (with weighing pan removed)
Pan boss is used (with the weighing pan and pan
support removed)
Pan support is used (with weighing pan removed)
Pan boss is used (with the weighing pan and pan
support removed)
Pan support is used (with weighing pan and pan
support plate removed)
Pan boss is used (with the weighing pan, pan
support plate and pan support removed)
Pan support is used (with weighing pan and pan
support plate removed)
Pan boss is used (with the weighing pan, pan
support plate and pan support removed)
Weighing
pan
Pan
support
Dust
guard
When the weighing pan and
pan support are removed
60 g
Mass range
110 g
to be weighed
110 g
60 g
Mass
of special
weighing pan
State
AD-4212A-600/1000
AD-4212A-200
When the weighing pan
is removed
Mass range
to be weighed
210 g
Mass
of special
20 g
weighing pan
86
Mass of
special weighing pan (g)
20 to 70
60 to 110
20 to 120
60 to 160
110 to 510
150 to 550
110 to 1000
150 to 1040
Pan support plate
When the weighing pan and
pan support are removed
110 g
Mass range
to be weighed
210 g
120 g
Mass
of special
60 g
weighing pan
110 g
160 g