Class
Item and Parameter
Cpout
Cp-t
Comparator
Comparator output
output
setting
Cp
Comparator mode
Cp-r *2
Comparison results
Cp fnc
Comparator
bep
LL buzzer
bep_
LO buzzer
bep-
OK buzzer
bep
HI buzzer
bep
HH buzzer
Cp HH
Secondary upper limit
Cp Hi
Upper limit
Cp lo
Lower limit
Cp ll
Secondary lower limit
dout
Data output
Not displayed
when BCD
prt
output (OP-01)
Data output mode
is installed.
*2 This item is not displayed for the AD-4212B series balance or when BCD output (OP-01) is installed.
*3 AD-4212A series balance: when OP-04 is installed
AD-4212B series balance: when "Cp-t" is set to "1"
0
3-level comparator contact output
1
5-level comparator contact output
0
No comparison
Comparison, excluding "near
1
zero" when stable or overloaded
Comparison, including "near
2
zero" when stable or overloaded
Continuous comparison,
3
excluding "near zero"
Continuous comparison,
4
including "near zero"
0
Not added
1
Added
0
OFF
1
ON
0
OFF
1
ON
0
OFF
1
ON
0
OFF
1
ON
0
OFF
1
ON
Refer to "10-10 Comparator
Function".
0
Key mode (when stable)
Auto print mode A
1
(Reference = zero)
Auto print mode B
2
(Reference = last stable value)
Stream mode /
3
Interval memory mode
4
Key mode B (instantly)
5
Key mode C (when stable)
Factory setting
30
Description
Only the AD-4212B series
balance displays this class and
item.
0: AD-4212B-23
1: Other models
Use in A&D standard format.
Displayed only when 5-level
*3
comparator output (
Displayed only when 5-level
*3
comparator output (
Displays Cp HH, , Cp ll only when
5-level comparator output (
selected.
Accepts the PRINT key or external
contact input to output (or store)
data only when the display is
stable.
Outputs (or stores) data when the
display is stable and conditions of
ap-p, ap-b and the reference value
are met.
With data 0, outputs data
continuously; with data 2, uses
interval memory.
Accepts the PRINT key or external
contact input to output (or store)
data, regardless of the display
condition.
Accepts the PRINT key or external
contact input to output (or store)
data, instantly when stable or after
the display is stabilized.
) is selected.
) is selected.
*3
) is