5 Confirm that there is nothing on the pan and press the
PRINT key. The balance measures the zero point and
displays the measured value. Do not allow vibration or
drafts to affect the balance.
The balance displays the calibration weight value.
6 Place a calibration weight, of the weight value displayed,
on the pan and press the PRINT key. The balance
measures the calibration weight and displays the
measured value. Do not allow vibration or drafts to affect
the balance.
7 The balance displays
pan.
8 The balance displays
Report" using the RS-232C interface or stores the
calibration test data in memory. For details on the
calibration test report format, refer to "11-2 GLP Report".
9 The balance will automatically return to the weighing
mode.
end . Remove the weight from the
glp
and outputs "Calibration Test
25
From previous page
Calibration
weight
GLP output