Thermo Scientific Orion Star User Manual page 76

Meter
Hide thumbs Also See for Orion Star:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Meter Setup Menu Features
Table of Conductivity Standard Values vs. Temperature
Cat. No.
111.9 mS/cm
Temperature
Conductivity
(°C)
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
EN-70
011005
011006
12.9 mS/cm
Conductivity
Standard
Standard
(mS/cm)
(mS/cm)
65.10
7.135
66.84
7.344
68.59
7.555
70.35
7.768
72.12
7.983
73.91
8.200
75.70
8.418
77.50
8.638
79.32
8.860
81.15
9.084
82.98
9.309
84.83
9.535
86.69
9.763
88.56
9.993
90.45
10.22
92.34
10.46
94.24
10.69
96.15
10.93
98.08
11.16
100.0
11.40
102.0
11.64
103.9
11.88
105.9
12.12
107.9
12.36
109.9
12.61
111.9
12.85
113.9
13.10
115.9
13.35
117.9
13.59
120.0
13.84
Thermo Scientific Orion Star
011007
01100910
1413 µS/cm
147 µS/cm
Conductivity
Conductivity
Standard
Standard
(µS/cm)
(µS/cm)
776
799
822
846
870
894
918
943
968
992
1017
1043
1068
1094
1119
1145
1171
1198
1224
1251
1277
1304
1331
1358
1386
1413
1441
1468
1496
1524
and Star Plus Meter User Guide
011008
100 µS/cm
Conductivity
Standard
(µS/cm)
81
54
83
56
86
58
88
59
91
61
93
63
96
64
98
66
101
68
103
70
106
72
108
73
111
75
114
77
116
79
119
81
122
83
125
85
127
87
130
88
133
90
136
92
138
94
141
96
144
98
147
100
150
102
153
104
156
106
159
108

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents