Thermo Scientific Orion Star User Manual page 63

Meter
Hide thumbs Also See for Orion Star:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Error Code
Description
202
Autosampler
is Jammed
203
Unstable Reading
from Autosampler
306
ISE Automatic
Blank Error
307
ISE Calibration
Standard Error
309
Bad ISE Slope
707
Conductivity
Calibration
Standard Error
709
Conductivity Cell
Constant Error
808
Bad Zero Point
DO Slope
809
Bad DO Slope
880
RDO
®
Probe Not Attached
881
RDO Optical
Probe Expired
882
RDO Optical
Probe Failure
Thermo Scientific Orion Star
Troubleshooting
Turn the autosampler off and wait 45 seconds before turning it back on.
The autosampler should return to the home position.
The measurements taken using the autosampler are unstable. Check the
electrodes for proper function. Make sure that the electrode cables are
properly connected.
Disable the automatic blank feature in the setup menu and re-calibrate
the meter without using a zero concentration standard.
The millivolts measured during calibration are the same for two
standards. Review the calibration procedure and verify that the electrode
was placed in the standards at the appropriate times. Clean the electrode
according to the electrode user guide. Re-calibrate the electrode with
fresh standards.
Clean the electrode according to the electrode user guide. Re-calibrate
the electrode with freshly prepared standards.
The conductivity value measured during calibration is the same for
two standards. Review the calibration procedure and verify that the
conductivity probe was placed in the standards at the appropriate times.
Clean the conductivity probe according to the probe user guide.
Re-calibrate the probe with new standards.
The cell constant is not in the range of 0.001 to 199.0 cm
conductivity probe according to the probe user guide. Re-calibrate the
probe with new standards.
An air calibration should be performed before the zero point calibration.
Verify that a solution with zero oxygen is being used for the zero point
calibration. A solution with 15 grams of Na2SO3 dissolved in 250 mL
of distilled water is recommended.
For polarographic DO probes, connect the probe to the meter, power on
the meter and let the probe to polarize for at least 30 minutes. For an air
calibration, check that the sponge in the calibration sleeve is damp and
there is no water on the probe membrane. For a water calibration, bubble
air into the sample and stir to keep bubbles off the membrane. Clean the
DO probe according to the probe user guide. Re-calibrate the DO probe.
Optical
Verify that the RDO probe is properly connected to the meter. Connect
another, known working RDO probe to the meter.
The optical cap attached to the RDO probe has expired. Install a new
optical cap according to the RDO probe user guide.
Verify that the optical cap on the RDO probe was properly installed,
not expired and not tampered with. Connect another, known working
RDO probe to the meter.
and Star Plus Meter User Guide
Troubleshooting
. Clean the
-1
EN-57

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents