A6 Interface Assembly Removal/Replacement

This module contains components that can be damaged or destroyed by
CAUTION
electrostatic discharge. It should be serviced at a static-safe
workstation. Refer to "Preparing a Static-Safe Work Station" in
Chapter 8.
To Remove the A6 Interface Assembly
1. Remove both covers. Unscrew one screw on each cover and then slide
2. Disconnect the ribbon cable from the A1 interconnect assembly to
3. Remove the rear panel from the main frame by removing the four
4. Remove the three screws that secure the A6 interface assembly.
To Replace the A6 Interface Assembly
1. Use the three screws to install the A6 interface assembly to the rear
2. Assemble the rear panel to the main frame using the four corner
3. Connect the ribbon cable from the rear panel to the A1 interconnect
4. Replace the module covers.
the cover toward the rear.
the rear panel.
corner screws.
panel.
screws.
assembly.