Cc Mode Test; Test Setup - Agilent Technologies 6060B Service Manual

300 watt single input electronic loads
Table of Contents

Advertisement

CC Mode Test

This test verifies that the Electronic Load operates in the CC Mode and that the current programming and readback to the
front panel display are within specifications. For each DMM reading, the front panel display should be equal to:
6060B: DMM reading in amps ± ((DMM reading in amps X 0.0005) + 0.065)
6063B: DMM reading in amps ± ((DMM reading in amps X 0.0012) + 0.010)
If the test readings significantly disagree with the specified values or no readings can be recorded, perform the CC MODE
TEST troubleshooting procedures in Figure 3-1 in Chapter 3. If the readings are out of tolerance, calibrate the applicable
current range (see Chapter 6 in the Operating Manual).
a.
Connect the Electronic Load, power supply (Agilent 6031A/6032A or equivalent), DMM, and the 0.010 ohm (6060B)
or 0.100 ohm (6063B) current monitor resistor as shown in Figure 2-1.
b. Turn on the Electronic Load.
c. Check the high amp current range as follows:
1. Press
MODE
6060B:
6063B:
2.
Turn on the power supply and set for:
6060B: 5V and >60A.
6063B: 5V and >10A.
3.
Wait 30 seconds and then record the DMM and front panel display readings. DMM reading should be between:
6060B: 598.7mV (59.865A) and 601.3mV (60.135A).
6063B: 997.5mV (9.975A) and 1.002 V (10.025A).
Note that the Electronic Load's CC annunicator is on.
4. Press
CURR
5.
Wait 30 seconds then record the DMM and front panel display readings. DMM reading should be between:
6060B: 9.24mV (0.924A) and 10.761mV (1.076A).
6063B: 98.85mV (0.9885A) and 101.15mV (1.0115A).
10
CURR
CURR
6
0
CURR
1
0
1
Enter
Figure 2-1. Test Setup A
Enter
, then
Enter
.
Enter
.
.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6063b

Table of Contents