Table 3-1: Analog Output Connections; Figure 3-5: Equipment Interface Connector Pin Arrangement - Teledyne 3010MA Operating Instructions Manual

Paramagnetic oxygen purity analyzer
Table of Contents

Advertisement

Part 1 Control Unit
are given where each function is described in the paragraphs below. At the
end of this section is a complete pin-out description.

Figure 3-5: Equipment Interface Connector Pin Arrangement

Analog Outputs: There are four dc output signal pins —two pins per
output. For polarity, see Table 3-1. The outputs are:
1. 0–1 V dc % of Range: Voltage rises linearly with increasing
oxygen, from 0 V at 95% O
scale is 100% O
2. 0–1 V dc Range ID: 0.25 V = Low Range, 0.5 V = Medium
Range, 0.75 V = High Range.
3. 4–20 mA dc % Range: (M option only) Current increases
linearly with increasing oxygen, from 4 mA at 0 ppm to 20 mA
at full scale. (Full scale = 100% O
4. 4–20 mA dc Range ID: (This option is not included unless
otherwise noted) 8 mA = Low Range, 12 mA = Medium
Range, 16 mA = High Range.

Table 3-1: Analog Output Connections

Pin
3
(+) Range ID, 4-20 mA, floating (special option)
4
(–) Range ID, 4-20 mA, floating (special option)
5
(+) % Range, 4-20 mA, floating
6
(–) % Range, 4-20 mA, floating
8
(+) Range ID, 0-1 V dc
23
(–) Range ID, 0-1 V dc, negative ground
24
(+) % Range, 0-1 V dc
7
(–) % Range, 0-1 V dc, negative ground
Alarm Relays: There are nine pins dedicated to the internal alarm
relay contacts. Each set of three pins provides one set of Form C relay
contacts. Each relay has both normally open and normally closed
Teledyne Analytical Instruments
to 1 V at full scale 100% O
2
for all ranges.)
2
Function
Model 3010MA
for all ranges.)
2
. (Full
2
30

Advertisement

Table of Contents
loading

Table of Contents