Table 3-1: Analog Output Connections; Table 3-2: Analog Concentration Output-Example - Teledyne 4060 Series Instruction, Operating And Maintenance Manual

Trace aromatic hydrocarbon analyzer
Table of Contents

Advertisement

Series 4060

Table 3-1: Analog Output Connections

Pin
Function
3
Channel 2, + 4-20 mA, floating
4
Channel 2, – 4-20 mA, floating
5
Channel 1, + 4-20 mA, floating
6
Channel 1, – 4-20 mA, floating
8
Channel 2, 0-1 VDC
23
Channel 2, 0-1 VDC, negative ground
24
Channel 1, 0-1 VDC
7
Channel 1, negative ground
Examples:
The analog output signal has a voltage which depends on gas
concentration relative to the full scale of the range. To relate the signal
output to the actual concentration, it is necessary to know what range the
instrument is currently on, especially when the analyzer is in the auto
ranging mode.
The signal output for concentration is linear over the currently
selected analysis range. For example, if the analyzer is set on a range
that was defined as 0-100 ppb benzene, then the output would be as
shown in Table 3-2.
To provide an indication of the range, the Range ID analog outputs
are used. They generate a steady preset voltage (or current when using
the current outputs) to represent a particular range. Table 3-3 gives the
range ID output for each analysis range.
Table 3-2: Analog Concentration Output—Example
Ppb
Voltage Signal
Benzene
Output (VDC)
0
0.0
10
0.1
20
0.2
30
0.3
Teledyne Analytical Instruments
Current Signal
Output (mA DC)
4.0
5.6
7.2
8.8
Installation
31