Manual Mode Spanning - Teledyne 3010MA Operating Instructions Manual

Paramagnetic oxygen purity analyzer
Table of Contents

Advertisement

Part 1 Control Unit
The beginning span value is shown in the upper left corner of the
display. As the span reading settles, the screen displays and updates
information on Slope. Spanning automatically ends when the slope is less
than 1/50 of the displayed value of the oxygen concentration (in ppm) for
three minutes. Then the instrument automatically returns to the analyze
mode.
4.4.1.2 M
ANUAL
Press Span to start the Span function. The screen that appears
allows you to select whether the span calibration is to be performed
automatically or manually.
keys to toggle between AUTO and MAN span settling. Stop
Use the
when MAN appears, blinking, on the display. Press Enter to move to the
next screen.
(<UP>) to permit modification (Mod #) of span value.
Press
Use the arrow keys to enter the oxygen concentration of the span gas
you are using. The < > arrows chose the digit, and the
the value of the digit.
Press Enter to enter the span value into the system and begin the
span calibration.
Once the span has begun, the microprocessor samples the output at a
predetermined rate. It calculates the difference between successive
samplings and displays this difference as
several seconds for the first
of change of the
S p a n
When the
S p a n
press Enter. (Generally, when the
the full scale of the range being calibrated for a period of five minutes it is
sufficiently stable.) Once Enter is pressed, the
the correct value. The instrument then automatically enters the Analyze
function.
Teledyne Analytical Instruments
M
S
ODE
PANNING
S p a n : S e t t l i n g : M A N
< E N T > F o r N e x t
S p a n V a l : 1 0 0 . 0 0
< E N T > S p a n < U P > M o d #
S l o p e
value to display.
S l o p e
reading. It is a sensitive indicator of stability.
# # # #
S l o p e = # # # #
value displayed on the screen is sufficiently stable,
reading changes by 1% or less of
S p a n
Model 3010MA
arrows choose
on the screen. It takes
indicates rate
S l o p e
%
S p a n
p p m / s
reading changes to
S p a n
52