Table 3-1: Analog Output Connections - Teledyne 3000ZA Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Installation

Table 3-1: Analog Output Connections

Pin
3
+ Range ID, 4-20 mA, floating
4
– Range ID, 4-20 mA, floating
5
+ ppm Range, 4-20 mA, floating (optional)
6
– ppm Range, 4-20 mA, floating (optional)
8
+ Range ID, 0-1 VDC
23
– Range ID, 0-1 V dc, negative ground
24
+ ppm Range, 0-1 VDC
7
– ppm Range, 0-1 VDC, negative ground
Alarm Relays: The nine alarm-circuit connector pins connect to the
internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 V ac into a
resistive load. The connectors are:
Threshold Alarm 1:
Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
Threshold Alarm 2:
Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
System Alarm:
Teledyne Analytical Instruments
Function
3000ZA
20

Advertisement

Table of Contents
loading

Table of Contents