Analog Output; Figure 3-2: Equipment Interface Connector Pin Arrangement; Table 3-1: Analog Output Connections - Teledyne 4020 Operating & Maintenance Manual

Total hydrocarbon analyzer
Table of Contents

Advertisement

Installation

Figure 3-2: Equipment Interface Connector Pin Arrangement

3.3.3.4 A
NALOG
There are four DC output signal pins—two pins per output. For
polarity, see Table 3-1. The outputs are:
0–1 VDC % of Range:
0–1 VDC Range ID:
4–20 mA DC % Range: Current increases linearly with
4–20 mA dc Range ID:

Table 3-1: Analog Output Connections

Pin
Function
3
+ Range ID, 4-20 mA, floating
4
– Range ID, 4-20 mA, floating
5
+ % Range, 4-20 mA, floating
6
– % Range, 4-20 mA, floating
8
+ Range ID, 0-1 VDC
Teledyne Analytical Instruments
O
UTPUT
Voltage rises linearly with increasing
concentration, from 0 V at 0
concentration to 1 V at full scale. (Full
scale = 100% of programmable range.)
0.20 V = Low Range
0.5 V = Medium Range
0.80 V = High Range
concentration , from 4 mA at 0
concentration to 20 mA at full scale.
(Full scale = 100% of programmable
range.)
6.8 mA = Range 1
12.0 mA = Range 2
16.8 mA = Range 3
Model 4020
18