System Self-Diagnostic Test - Teledyne 3010MA Operating Instructions Manual

Paramagnetic oxygen purity analyzer
Table of Contents

Advertisement

Paramagnetic O2 Purity Analyzer

4.3.5 System Self-Diagnostic Test

The Model 3010MA has a built-in self-diagnostic testing routine.
Pre-programmed signals are sent through the power supply, output board
and sensor circuit. The return signal is analyzed, and at the end of the test
the status of each function is displayed on the screen, either as OK or as a
number between 1 and 3. (See System Self Diagnostic Test in Chapter 5
for number code.)
The self diagnostics are run automatically by the analyzer whenever
the instrument is turned on, but the test can also be run by the operator at
will. To initiate a self diagnostic test during operation:
Press the System button to start the System function.
Use the < > arrow keys to blink More, then press Enter.
Use the < > arrow keys again to move the blinking to the Self–Test
function. The screen will follow the running of the diagnostic.
During preamp testing there is a countdown in the lower right corner
of the screen. When the testing is complete, the results are displayed.
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting and Table 5-1 for number-code
information. The results screen alternates for a time with:
Then the analyzer returns to the initial System screen.
Teledyne Analytical Instruments
C o n t r a s t A u t o — C a l
P S W D L o g o u t M o r e
V e r s i o n S e l f — T e s t
S e n s o r O u t p u t : # # #
R U N N I N G D I A G N O S T I C
T e s t i n g P r e a m p — 8 3
P o w e r : O K
A n a l o g : O K
P r e a m p : 0
P r e s s A n y K e y
T o C o n t i n u e . . .
Operation
49