The Range Function - Teledyne 3000ZA2G Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Operation
Once the parameters for alarm 1 have been set, press Alarms again,
and repeat this procedure for alarm 2 (AL–2).
To reset a latched alarm, go to Dft– and then press either
two times. (Toggle it to Y and then back to N. )
times or
V
Go to Ltch– and then press either ▲ two times or ▼ two times.
(Toggle it to N and back to Y. )

4.6 The Range Function

The Range function allows the operator to program up to three
concentration ranges to correlate with the DC analog outputs. If no
ranges are defined by the user, the instrument defaults to:
The Model 3000ZA2G is set at the factory to default to auto
ranging. In this mode, the microprocessor automatically responds to
concentration changes by switching ranges for optimum readout
sensitivity. If the current range limits are exceeded, the instrument will
automatically shift to the next higher range. If the concentration falls to
below 85% of full scale of the next lower range, the instrument will
switch to that range. A corresponding shift in the DC percent-of-range
output, and in the range ID outputs, will be noticed.
The auto ranging feature can be overridden so that analog output
stays on a fixed range regardless of the oxygen concentration detected. If
the concentration exceeds the upper limit of the range, the DC output
will saturate at 1 VDC (20 mA for the current output).
However, the digital readout and the RS-232 output of the
concentration are unaffected by the fixed range. They continue to read
accurately with full precision. See Front Panel description in Chapter 1.
The automatic air calibration range is always 0-25 % and is not pro-
grammable.
Teledyne Analytical Instruments
–OR –
Low = 0–10 ppm
Med = 0–100 ppm
High = 0–1000 ppm
3000ZA2G
two
A
48

Advertisement

Table of Contents
loading

Table of Contents