YOKOGAWA TM20 User Manual page 90

Thermo-collector
Hide thumbs Also See for TM20:
Table of Contents

Advertisement

7 Other Useful Functions
6.
Press the [SET] key.
[TIP]
It is also possible to modify the measurement interval setting before pressing the [SET] key.
7.
Press the [SET] key.
The measurement period (the number of days from the options which also include the number of hours, minutes, and
seconds) is displayed.
[TIP]
It is also possible to modify the number of days setting before pressing the [SET] key.
8.
Press the [SET] key.
The measurement period (the number of hours, minutes, and seconds from the options which also includes the
number of days) is displayed.
[TIP]
It is also possible to modify the number of hours, minutes, and seconds settings before pressing the
[SET] key.
9.
Set the date (year/month/day) and time when logging starts.
10.
Press the [SET] key.
The confirmation screen appears.
11.
Press the [1] key.
The
mark and"Stand by" is displayed in the screen, and the instrument will enter the wait state. When the time
set in step 9 is reached, the measurement will automatically begin.
21:30:20
S t a n d by
[TIP]
To cancel the wait state, hold down the [ESC] key for about two seconds.
IM 54011-E
-End of procedure-
81
81